principle

The acid-base concentration meter is mainly used to monitor the concentration of the regeneration liquid of the ion exchanger in the process of producing high-purity water by the ion exchange method, or to prepare the pickling liquid for boilers and pipelines, and to continuously monitor the concentration of hydrochloric acid in the solution.

The AE30 type concentration meter adopts the electrode principle and is suitable for the AE30B polysulfone platinum electrode sensor, which can measure the concentration of various media such as HC, NaOH, H2SO4NaCl, HNONa2CO3, etc.

The concentration of the solution is indirectly calculated by measuring the conductivity of the solution. It is known that at a constant temperature, the conductivity of the low-concentration electrolyte corresponds to the concentration of the solution. When the concentration remains unchanged but the solution temperature changes, Conductivity also changes, ie the concentration of the solution is a function of conductivity and temperature. If the temperature of the solution can be measured and corrected to the conductivity at the standard temperature according to the aforementioned corresponding relationship, it can be directly converted into the concentration of the solution. It should be emphasized that the concentration of the solution is the sum of the conductivity of all conductive ions in the water. If there are two or more conductive components in the water, the above concentration-temperature-conductivity data can be measured, and the instrument can also be used. Measurement. If the ratio of the conductive composition in the water is uncertain, the meter cannot be used to measure.

In order to avoid electrode polarization, the instrument generates a square wave signal with high stability and applies it to the conductivity cell. The current flowing through the conductivity cell is proportional to the concentration of the measured solution. After the secondary meter converts the current from a high impedance operational amplifier to a voltage, After the program-controlled signal is amplified, detected and filtered, the potential signal reflecting the concentration is obtained. The microprocessor samples the temperature signal and the conductivity signal alternately, and after calculation and temperature compensation, the concentration value at a constant temperature is obtained.

application

Chemical, sewage, water supply, power generation, metallurgy, biology, pharmaceutical, paper and other industrial processes

Features

● Automatic temperature compensation, more precise measurement data

● Phase-sensitive detection: eliminate the influence of the wire on the concentration measurement

● Automatic conversion of measurement frequency: avoid electrode polarization and improve measurement accuracy

● Display concentration, output current, temperature, time and status on the same screen

● Adjustable uniform background light, can clearly observe the operation in the dark outdoor conditions

● Quick response, stable work and high precision

● Photoelectric coupling isolation output, good electromagnetic compatibility, high reliability, strong anti-interference ability

● Manual current source function: the output current value can be checked and arbitrarily set to facilitate the detection of the recorder and the lower computer

● Chinese display, Chinese menu

technical parameter

Measuring range:HCI(0.00~10.00)%Output current error≤±1%FS
NaOH(0.00~10.00)%Temperature coefficient compensation range(0~10.0)%/C (adjustable)
H2SO4(0.00~30.00)%Alarm relayAC220V 3A
NaCI(0.00~20.00)%Communication InterfaceRS485
HNO3(0.00~20.00)%power supplyAC220V±22V, 50Hz±1Hz;
Na2CO3(0.00~10.00)%DC24V±2.4V, 1A power supply
Temperature Compensation Range0~99.9°CProtection classIP65
Electronic unit basic errorConcentration: ±1.0%,data storage1 month (1:00/5:00)
Temperature: ±0.3°Cambient temperature0~60°C
Instrument basic errorConcentration: ±2.0% (the error can be less than 0.05% after calibration of common points)Relative humidity<85%
Temperature: ±0.5°C (0.0°C CsT≤60.0°C)Sensor materialPolysulfone.PSF, Platinum
Electronic unit stability±1.0%/24hsensor operating temperature0~99.9°C
Galvanically isolated output4~20mA(load<750Ω)sensor working pressure≤1.0Mpa

Selection table

AE30A-Online electrode type acid-base concentration meter transmitter (wall-mounted)
AE30B-Online electrode type acid-base concentration transmitter (pan mounted 138*138)

Measuring range

HCI (0.00...10.00)%

NaOH (0.00..10.00)%

H2SO4 (0.00...30.00)%

NaCl (0.00...20.00)%

HNO3 (0.00...20.00)%

Na2CO3 (0.00...10.00)%

power supply

AC220V

DC24V

output and communication


4...20mA current signal

4...20mA current signal + high and low alarm switch

(special option) VARIO

no special options

RS485 communication

Matching IB10 ABS material instrument protection box

Matching IB10 1.4301 (304) material instrument protection box (indoor)

Matching IB10 1.4435 (316L) material instrument protection box (outdoor)

Matching V10 sunshade
The above selection table does not contain selection codes. For selection, please contact the sales staff or email to sales@kewill-auto.cn to obtain a complete sample.
Prev :No
Next:AE50 Series Online Electromagnetic Acid-Base Concentration Meter
主站蜘蛛池模板: 日本高清不卡一区久久精品 | 淫五月天 | 善良的嫂子3中文 | 99成人综合久久精品亚洲直播秀 | 五月婷婷综合激情网 | 五月婷婷六月激情 | 国产高清不卡一区二区 | 日韩欧美视频在线播放 | 九九视频这里只有精品 | 国产日韩欧美视频在线 | 亚洲激情综合 | 久久精品国产99精品最新 | 成人小视频网 | 玖玖五月| 国产小视频免费看 | 开心色99xxxx开心色 | 免费www视频 | 国产国语一级毛片 | 日韩a一级欧美一级 | 免费观看120秒 | 深爱五月激情五月 | 国产成人在线精品 | 九九精品国产兔费观看久久 | 精品免费久久 | 久久免费观看视频 | 高清女主播一区二区三区 | 加勒比一区二区 | 男女羞羞视频免费观看 | 免费黄色国产视频 | 日韩在线视 | 国产亚洲福利精品一区二区 | 色五月天天 | 男女网站视频 | 亚洲国产综合视频 | 99久久香蕉 | 久久青草18免费观看网站 | 色视频国产 | 日韩第十页| 丁香花婷婷 | 国产精品久久久久久久久久久久久久 | 欧美洲精品亚洲精品中文字幕 |